REMSCAN

QUICK START

INSTRUCTIONS

 

PDA v1.12

PREPARATION PRIOR TO SCANNING SAMPLES

1.  In advance, ensure that the batteries in the Personal Digital Assistant (PDA) and the instrument are 

fully charged and loaded as per the Operations Manual instructions. The PDA battery will give up 

to 8 hours of regular scanning and the instrument batteries will each give at least 3 hours of regular 

scanning. It is good practice to ensure that one instrument battery is charging while the other is in 

use to allow continuous scanning.

2.  Turn on the RemScan instrument using the 

button at the base of the handle below the trigger

. The 

button will flash red/green for a few seconds before turning solid green. When the light is green the

 

instrument is turned on. It is recommended that the instrument is turned on BEFORE the PDA is 

turned on.

3.  Start the PDA as follows:
 

a.  Once the instrument is turned on, using the stylus (or a pen) press the   at the top right hand 

corner of the PDA once and release (see diagram over page). The green Windows Mobile 

home screen will appear.

 b.

  Push the “START” button on the PDA front panel (see diagram over page).

4.  After a few seconds a warm up screen will appear with a blue/green progress bar in the centre 

of the screen. The message 

“Device is warming up”

 will appear. After around 15 minutes the 

instrument will be warmed up and a message 

“Background Required. Please attach Background 

cap and press trigger”

 will appear.

5.  Place the Background Cap (black cap labeled “Background”) on the nose cone of the instrument 

and ensure it is secured and clipped on correctly. Pull the trigger and a blue/green progress bar will 

appear and the scan will take around 1 minute to complete.

6.  If the Background Cap scan was unsuccessful the message 

“Background energy is out of range. 

Make sure the background cap is clean and firmly in place”

 or 

“Background Failed”

 will appear. 

Please ensure that the Background Cap and Nose Cone is clean and dry. This can be done using 

deionized water but MUST be thoroughly air-dried before using. If background material is noticably 

dirty, replace with spare Background cap supplied with kit. Then ensure the cap is attached firmly 

and follow the prompts on the screen to repeat the calibration process.

7.  If the Background Cap scan was successful, the message 

“Background Scan Complete. 

Reference Scan Required”

 will appear. Place the Reference Cap (white cap labeled “Reference P/S”) 

on the nose cone and ensure it is secured and clipped on correctly. Pull the trigger and a blue/green

 

progress bar will appear and the scan will take around 1 minute to complete.

8.  If the reference scan is unsuccessful the message

 “Re-present sample” 

or 

“Reference Failed” 

or 

“Reference energy is out of range. Make sure the reference cap is clean and firmly in place”

 will be 

displayed on the screen. Please ensure that the Reference Cap is clean and dry and that the cap 

is attached firmly and follow the prompts on the screen to repeat the calibration process.

9.  If the Reference Cap scan is successful the instrument is now ready for scanning soil and a 

message will be displayed 

“instrument is now ready for scanning soil”

.

10.  If the Background and Reference Caps are both clean and dry and the calibration steps are 

unsuccessful after several attempts, please contact Ziltek for technical assistance.

11.  The PDA will ask the user to repeat the background standarization (steps 5-8) every hour.
 

NOTE: A quick reference check may be required to calibrate the instrument before the next 

background is needed. A message will be displayed saying that a quick reference check is  

required. Clean the nose cone and place the Reference P/S cap on and click the trigger.

12.  Select Calibration Model - Press the “Admin” icon and enter password, the default password is 

“admin”, then click “Login”. Select desired calibration model from the drop down menu, click “Save”.

W ziltek.com

Z050-14  07/18

TECHNICAL ASSISTANCE

For technical assistance, please contact Ziltek on [email protected]