           OPERATING      INSTRUCTION 

                            ZI-9940  

                 AC CLAMP METER

 

 

               International Safety Symbols 

This symbol, adjacent to another symbol or terminal, indicates the user must refer to 
the manual for further information. 

 

This symbol, adjacent to a terminal, indicates that, under normal use, hazardous 
voltages may be present 

 

Double insulation 

 

SAFETY NOTES 

  Do not exceed the maximum allowable input range of any function 

  Do not apply voltage to meter when resistance function is selected. 

  Set the function switch OFF when the meter is not in use. 

WARNINGS

 

  Set function switch to the appropriate position before measuring. 

  When measuring volts do not switch to current/resistance modes. 

  When changing ranges using the selector switch always disconnect the test leads from the 

circuit under test.

 

  Do not exceed the maximum rated input limits.

 

CAUTIONS

 

Improper use of this meter can cause damage, shock, injury or death. Read and understand 
this user manual before operating the meter. 
Always remove the test leads before replacing the battery. 
Inspect the condition of the test leads and the meter itself for any damage before operating the 
meter. Repair or replace any damage before use. 
Use great care when making measurements if the voltages are greater than 25VAC rms or 
35VDC. These voltages are considered a shock hazard. 
Remove the battery if the meter is to be stored for long periods. 
Always discharge capacitors and remove power from the device under test before performing 
Diode, Resistance or Continuity tests. 

 Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty 

of connection to the recessed electrical contacts. Other means should be used to ensure that 
the terminals are not "live". 

 If the equipment is used in a manner not specified by the manufacturer, the protection 

provided by the equipment may be impaired. 

Input Limits 

Function 

Maximum Input 

A AC 

200A 

V DC, V AC 

600V DC/AC 

Resistance, Diode, Continuity Test 

600V DC/AC 

 

Meter Description 

1.  Current clamp 
2.  Clamp trigger 
3.  Rotary Function swith 
4.  LCD display 
5.  MAX Hold/ Backlight button 
6.  Mode select button 

7. 

COM input jack  

8. 

Ω

 jack  

9. 

Battery Cover 

  

1. 

AC  DC 

     AC (alternating current) and DC (direct currrent) 

2. 

                  Minus sign 

3. 

1.8.8.8

 

  2000 count (0 to 1999) measurement reading 

4. 

AUTO 

      AutoRange mode 

5. 

MAX 

        MAX Hold mode 

6. 

                 Diode test mode  

7. 

   

•)))

        Audible Continuity 

8. 

HOLD 

     Data Hold mode  

9.      

BAT 

      Low Battery icon      

 

10.

 m,V,A,K,M,

Ω

  Units of measure list 

Specifications 

Function 

Range & Resolution 

Accuracy (% of reading) 

AC Current 

200.0 AAC 

± (2.5% + 10

 digits

)

 

DC Voltage 

200.0 mVDC 

± 

(

0.5% + 5 digits) 

2.000 VDC 

± (1.2% + 3 digits) 

20.00 VDC 
200.0 VDC 
600 VDC 

± (1.5% + 3 digits) 

AC Voltage 

2.000 VAC 

± (1.5% + 3 digits) 

20.00 VAC 
200.0 VAC 
600 VAC 

± (2.0% + 4 digits) 

Resistance 

200.0 

 

± (1.0% + 4 digits) 

2.000K

 

± (1.5% + 2 digits) 

20.00K

 

200.0K

 

2.000M

 

± (2.0% + 3 digits) 

20.00M

 

± (3.0% + 5 digits) 

 

Clamp size                           

Opening 0.7" (18mm) approx 

Diode Test

 

                   

Test  current of 0.3mA typical; Open circuit voltage 1.5V DC 

typical. 

Continuity Check

 

      Threshold <120

; Test current < 1mA 

Low Battery Indication

 

   

“ BAT” is displayed 

Overrange Indication 

   

“OL  “ is displayed 

Measurements Rate          

2 per second, nominal 

Input Impedance

 

       7.8M

 (VDC and VAC) 

Display 

3-1/2 digits (2000 counts) LCD 

kM 

BAT 