Type BM24L10 EHEDG Load Cell
Specifications and dimensions are subject to change without notice and do not constitute any liability whatsoever.
Zemic Europe B.V.
T: +31 76 50 39480
Nr. 2023.11 BM24L10 Installation and Operating Manual Rev1
Page 1/7
1 Safety Instructions
1.1
Specific Condition of Use
The BM24L10 series
1
weighing sensors are suitable for technical weighing applications within the
specified environment and measuring range, and are not suitable for any purpose beyond
the specified range.
The sensor has to be installed by qualified personnel in accordance with safety specifications,
technical requirements and these installation instructions.
In addition, please properly transport, store, install and operate with care according to the
installation and cleaning instructions.
1.2 Safety Loading
The load cell must be operated in accordance with the technical data sheet, in particular, the
specified safety overload requirement.
It must be noted that when multiple load cells are mounted on single scale, the load on
every single load cell is not always evenly distributed and care must be taken not to overload them.
1.3 Used as Equipment Unit
The load cell can be used as an equipment unit. However, it must be noted that in order to improve
the output sensitivity, the design of the load cell does not conform to the common safety factors in
mechanical design. Please refer to the "Safety Loading" and "Technical Specifications" for details.
1.4 Additional Safety Precautions
As a passive component, when the equipment fails, the sensor does not have any ability to actively
cut off the fault, nor does it perform any obligation to actively cut off the fault.
This function needs to be considered and designed by the users themselves.
If the load cell fails or breaks and may cause damage to the equipment or casualties, the user must
take appropriate safety measures, such as automatic emergency shutdown, overload protection,
and other anti-fall protection, etc.
1
‘
series’ refers to the fact that this model load cell
is available in three different capacities with the same outer dimensions.