Installation
Notes on installation and use
WARNING - GENERAL HAZARDS
Do not store or operate the device in ambient conditions other than those
prescribed.
The device should be set up so that the power cable can be disconnected from
the power supply quickly and easily without any tools.
WARNING - RISK OF ELECTRIC SHOCK
When connecting the device, the operator is responsible for operating of the
device and for meeting the safety and EMC requirements defined in
IEC 60601-1, chapter 16 (medical electrical systems) and IEC 60601-1-2.
Additional extension cables or portable multiple sockets may not be used.
The electrical installation must conform to IEC 60364-7-710 or the applicable
national regulations. This includes the existence of a ground fault circuit
interrupter (GFCI).
To avert the risk of an electric shock this device may only be connected to a
power supply with a protective earth conductor.
WARNING - FIRE HAZARD
The device is not suitable for operation in explosion risk areas (e.g.
combustible mixture of anesthetic, cleaning or disinfecting agents with air,
oxygen or nitrous oxide).
The electrical installation must conform to IEC 60364-7-710 for USA and
CANADA only: For single-phase power systems with 120 V AC, NEMA 5-
15P power plugs must be used.
CAUTION –PROPERTY DAMAGE
Do not store or use this device in damp rooms. Do not expose the device to
water splashes, dripping water or sprayed water.