9 Maintenance
Documentation set
VISULAS green
9.5 Replacing the light-emitting diode (LED) of the fixation lamp (optional)
90 / 116
000000-2096-101-DokS-GB-181218
The fuse compartment with the fuse cartridge and the two fuses is
located at the right-hand side of the power input of the laser
console.
1
2
3
Figure 42: Replacing the fuses
1
Fuse compartment
2
Fuse
3
Fuse carrier
Action
1. Remove the fuse carrier with fuse by pressing gently on the
snap-in clip.
2. Replace the defective fuse.
3. Re-insert the fuse carrier with fuse. Ensure that the fuse carrier
snap-in clip is correctly positioned.
9.5 Replacing the light-emitting diode (LED) of the
fixation lamp (optional)
NOTE
Damage by unsuitable LED
Installing an unsuitable LED may cause the fixation lamp to
overheat and be destroyed.
u
Operate the optional fixation lamp of the laser slit lamp only
with the standard red-flashing LED as supplied.