![Zeiss Smartzoom 5 Скачать руководство пользователя страница 6](http://html2.mh-extra.com/html/zeiss/smartzoom-5/smartzoom-5_quick-manual_4422524006.webp)
3 Referencing Tasks | 3.1 Stage (Angle-Stops)
ZEISS
3 Referencing Tasks
This chapter describes how to perform the referencing tasks available. We recommend to perform
the full referencing workflow (meaning all referencing tasks are activated) if the system was newly
installed. The entire workflow will take about 60 minutes with all objectives. When you exchange
a component (e.g. the optical engine) we recommend to perform the full referencing workflow
again.
3.1 Stage (Angle-Stops)
This referencing task will find the best settings for the microscope stage. Therefore the angle-
stops of the stage insert plate must be facing up. As you have to mount an objective, please read
the Smartzoom 5 instruction manual, where this procedure is described in detail. Note that you
should always start referencing with having mounted the objective with the highest magnification
(this depends on which objective you have bought).
1. Mount the objective with the highest magnification (e.g.
10.0x
objective). Ensure that the
stage is empty and that the stand is in upright position. The angle-stops of the stage insert
plate must be facing
up
.
2. Click on the
Start referencing
button in the Referencing Tasks Area.
à
The
Referencing Wizard
will appear in the preview area.