4 Troubleshooting | 4.3 Stage does not move to the displayed position
ZEISS
4. If stand is upright:
- Cancel the referencing (with saving)
- Select
Tilting Accelerometer
and all not referenced tasks
- Start the referencing task(s)
4.3 Stage does not move to the displayed position
In rare cases during image alignment, stage or objective referencing it can happen that the stage
moves more than 10 mm besides the target position.
1. Cancel the current referencing and close the software
2. Switch the microscope off remove the power cables for at least 1 minute
3. Reconnect the power cables and start the microscope
4. Start the referencing app and perform the referencing again. If the red cross on the micro-
scope appears after the restart or the problem persists after the restart, contact the ZEISS
customer support.
4.4 Stitching Referencing: Referencing Object not found
The error "
Referencing Object not found
” appears during the Stitching Referencing.
1. Check if the right alignment sheet is used.
à
The new alignment sheet (
2294-728
, rear) is the correct one, not the old alignment
sheet (2115-856).
4.5 Best Practice: Image Alignment
In order to get the best possible overview of the image ratio, the following tips should be fol-
lowed.
1. Selection of the objective: We recommend to take 5,0x or 1,6x if possible.
Quick Guide ZEISS Smartzoom 5 | en-US | Rev.1
28