ZEISS
6 Care and Maintenance | 6.4 Care and Cleaning Work
6.3.4 Firmware Update of the Primostar 3 HD IP Camera
Prerequisite
ü
The Primostar 3 HD IP camera is connected to a PC or laptop.
ü
USB flash drive is available (included in the scope of delivery)
Procedure
1. Download the latest firmware from the website.
2. Save the update file to the root folder of the USB flash drive.
3. Insert the flash drive into the Primostar 3 HD IP camera's USB port or a connected USB hub.
4. In the
OSD
menu, click
Settings > Operating System Settings > Firmware Update
.
à
The firmware is updated automatically.
6.4 Care and Cleaning Work
DANGER
Electric shock due to live parts
When the Microscope is still switched on, coming in contact with live parts can lead to electric
shock or burn.
4
Switch off Microscope prior to opening or cleaning.
4
Disconnect live parts from the power supply.
NOTICE
Functional impairment due to dirt and moisture
Dirt, dust and moisture can impair the Microscope’s functionality and can cause short-circuits.
4
Use the dust protection cover if the Microscope is not used.
4
The ventilation slots must be unobstructed at all times.
4
Perform regular maintenance and cleaning according to the instructions in this document
and according to the instructions in the applicable documents.
4
Make sure that no cleaning liquid or moisture gets inside the Microscope.
4
In case of damage, the affected parts of the Microscope must be taken out of operation.
6.4.1 Cleaning the Primostar 3
Auxiliary materials
Quantity
Optics cleaning solution, consisting of 90 vol% gasoline and 10 vol% iso-
propanol (IPA).
Its individual constituents are also known as:
Gasoline: Surgical spirit, petroleum ether
Isopropanol: 2-propanol, dimethyl carbinol, 2-hydroxypropane
as needed
Tab. 9: Auxiliary materials required for care
Prerequisite
ü
Microscope is switched off and disconnected from power supply.
Procedure
1. Cover the instrument with the dust cover after every use.
2. Do not install the microscope in a humid room; maximum humidity is 75 %.
3. Cover open tubes with the dust caps.
4. Remove dust and loose dirt from visible optical surfaces with a brush, blower brush, cotton
bud, optics cleaning tissue, or a cotton cloth.