4 Installation | 4.3 Installing Additional Components
4.3.11 Installing/Removing the Mirror
The mirror serves to illuminate the specimen if no power outlet is available.
Parts and Tools
Hex Key
Prerequisite
ü
Fixed-Köhler stand or Full-Köhler stand
Procedure
1. Remove the cover cap from the luminous-field diaphragm, see
low Filter or TL Cover Plate (Fixed Köhler Stand) [
.
Installing/Removing the Condenser [
.
3. Unscrew the 2 clamping screws from the con-
denser carrier insert.
4. Push the condenser carrier insert backward against the spring, and inclining it, remove it
upward from the condenser carrier.
5. Insert the mirror from the top through the
opening of the condenser carrier and place it
onto the mount of the luminous-field di-
aphragm. Take care that the mirror rests hori-
zontally on the mount.
6. Rotate and incline the mirror until the daylight is reflected homogeneously into the light
path.
Info
Follow these instructions in reverse order to remove the mirror.
4.3.12 Inserting an Eyepiece Pointer or an Eyepiece Micrometer
Parts and Tools
Hex Key, 1.0 mm
Procedure
1. Use an Allen wrench to loosen the set screw on
the binocular body from below.
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en