Preparation for use
OPMI Lumera i
Version 8.0
Page 80
G-30-1720-en
Setting up the system
Limiting the suspension arm’s downward movement
Proceed as follows to limit downward movement:
•
Loosen locking lever (1) for the upward/downward movement.
•
Lower the surgical microscope until it can be focused on the surgical field
(depending on the focal length of the objective lens). At the same time,
ensure that there is sufficient safety distance to the surgical field.
•
Firmly retighten locking lever (1) for the upward/downward movement.
•
Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
The limit of downward travel is only effective in the lower part of the suspension
arm’s vertical movement range (starting from the arm’s horizontal position).
Fig. 30: Adjusting the
downward travel limit
CAUTION
Risk of injury caused by lowering of the surgical microscope!
If the surgical microscope has not been correctly balanced, it may lead to
downward movement and cause injury to the patient.
•
Adjust the limit of downward travel of the suspension arm in such a way
–
that the surgical microscope cannot come into contact with the
patient even if it is inadvertently lowered.
–
that during coarse focus with the suspension arm there is sufficient
distance maintained from the patient, particular if a fundus viewing
system and other accessories are mounted.
1
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...