Version 8.0
G-30-1720-en
Page 143
OPMI Lumera i
Care and maintenance
Sterilization
The asepsis sets available from ZEISS contain rubber caps and hand grips that
can be sterilized in autoclaves. For detailed information on sterilization, please
see the enclosed instructions “Preparation of resterilizable products” for the
respective asepsis set.
You can also use disposable sterile drapes to cover the system.
•
Loosely attach the sterile drapes so that there is sufficient space for the
microscope supporting stand and the surgical microscope to move.
•
Do not cover the ventilation slots in order to ensure that the lamps are
able to cool sufficiently and to avoid any loss in lamp functionality.
Disinfection
It may be necessary to disinfect the device surfaces.
The maximum application concentrations are:
–
For alcohol (tested with 2-propanol): 60%
–
For aldehyde (tested with glutaraldehyde): 2%
–
For quaternary compounds (tested with DDAC): 0.2%
NOTE
Surface damage caused by wrong disinfectants!
Performing disinfection with the wrong disinfectants may result in damage
to the surfaces of the device.
•
Use an aldehyde and/or alcohol-based disinfectant. The addition of
quaternary compounds is acceptable.
•
In order to prevent surface tensions, you may use only the disinfecting
components specified below.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...