G-30-1720-en
Adjusting the eyepieces
–
Emmetropes:
•
Set the diopter setting ring (5) to 0 D. Set (diopters).
–
Eyeglass wearers who perform surgery wearing their eyeglasses:
•
Set the diopter setting ring (5) to 0 D.
–
Eyeglass wearers who perform surgery without their eyeglasses (prescrip-
tion known):
•
Set the diopter setting ring (5) to the respective value.
–
Eyeglass wearers who perform surgery without their eyeglasses (prescrip-
tion not known):
•
Set the diopter setting ring (5) on the main microscope to + 5 D.
•
Set the diopter setting ring of the assistant’s microscope, if provided,
to + 5 D.
•
Adjust the eyecups (4) in such a way that the entire field of view can
be seen.
Adjusting the magnification
•
Set the maximum magnification on the main microscope using the foot
control panel.
•
Focus on the object.
•
Set the required magnification using the foot control panel. The focal
plane is retained.
Eyepieces without reticle
Eyepieces with reticle
•
Remove the tube with the
eyepieces from the microscope
body and point it at a distant
object.
•
Slowly turn the diopter setting
ring (5) of one eyepiece clockwise
until the object is in focus.
•
Mount the tube with eyepieces
on the microscope body and firm-
ly tighten the securing screw (3).
•
Slowly turn the eyepieces on
the diopter setting ring (5)
clockwise until the reticle is
in focus.
•
Focus the main microscope on
the object. Reticle and object
must be in focus at the same
time.
–
Observation with
eyeglasses:
Screw in the eyecups (4) all the way.
–
Observation without
eyeglasses:
Screw out the eyecups (4) and adjust
them for your field of view.
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...