OPMI Lumera
®
i on floor stand
Safety measures
Version 9.2
G-30-1720-en
Page 17
Transport & service
•
The system may be transported over long distances (e.g. during reloca-
tion, return for repair, etc.) only in its original packaging or in special
return packaging. Please contact your dealer or ZEISS Service for this
purpose.
•
If there is a malfunction, and you cannot resolve the issue by referring to
the section “What to do in the event of malfunctions”, attach an “Out of
order” sign to the device and contact ZEISS Service.
Safety inspection
•
To prevent any impairment of the system’s safety as a result of ageing,
wear, etc., the operator must ensure, in accordance with the applicable
national regulations, that the regular technical safety checks defined for
this system are performed on schedule and to the stipulated extent.
The technical safety checks must only be performed by ZEISS or persons
authorized by ZEISS. The following safety inspections should be
performed as a minimum for the system:
–
Availability of the instructions for use
–
Visual inspection of the device and its accessories for damage, as well
as legibility of markings and labels
–
Leakage current test
–
Test of protective ground conductor
–
Function and wear test of the brakes
–
Functional check of all switches, buttons, connectors, and indicator
lamps on the device
–
Functional test of the manual mode without the patient, at least every
six months
Medical electrical systems
Any additional equipment you connect to medical electrical devices
must demonstrably comply with the applicable standards and directives
(e.g. IEC 60950-1 for data processing equipment).
In addition, all configurations must meet the standard requirements for
medical electrical systems (see IEC 60601-1).
Anyone connecting additional equipment to medical electrical devices is a
system configurer and, as such, is responsible for compliance of the system
with the normative requirements for medical electrical systems.
Please note that local legislation takes priority over the above-mentioned
normative requirements. If you have any questions, contact your local dealer
or ZEISS Service.
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...