Device data
OPMI Lumera
®
i on floor stand
Version 9.2
Page 140
G-30-1720-en
Technical data
Essential performance features OPMI Lumera i
OPMI Lumera i does not have any essential performance features as defined
in IEC 60601-1.
Instead of the device’s essential performance features, certain device func-
tions defined by the manufacturer have been tested.
The following device functions have been defined for the immunity test:
–
Visualization of the object must be maintained (minor image distortions
are permitted).
–
The lighting for object illumination must be maintained (interruptions of
less than one minute are permitted).
–
The zoom must be adjustable.
–
Focusing must be possible.
–
Operation via the foot control panel must be maintained.
–
Operation via the control panel must be maintained.
–
The video image must be available.
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...