10 Troubleshooting
Instructions for Use
OPMI LUMERA 700
10.4 Troubleshooting work
190 / 236
G-30-1673-en - 14.0 - 2018-07-12
10.4 Troubleshooting work
10.4.1 Activating manual mode
1
Figure 77: Activating manual mode
1
[Manual mode] switch
CAUTION!
Risk of injury caused by malfunctioning motor electronics!
In the case of malfunctioning motor electronics, the device may
make uncontrollable movements or main functions such as XY
movement, focus, zoom and light control may become impaired.
u
Activate manual mode.
u
Contact ZEISS Service.
Action
1. Press the [Manual mode] switch.
ð
The light sources are at medium intensity
ð
The XY coupling, focus and zoom motor are inactive
ð
The filters are the in swiveled-out position. For filters that
were previously in the swiveled-in position: note the
"Maximum radiation exposure times" table. [
ð
The control panel is black
ð
The shutters are in the swiveled-out position
2. If manual mode is activated: operate the device manually.