6 Before every use
Instructions for Use
OPMI LUMERA 700
6.3 Function test
126 / 236
G-30-1673-en - 14.0 - 2018-07-12
ð
For xenon lamps: the remaining service hours must be suffi-
cient to complete the operation.
u
Check to make sure the objective lens and the eyepieces are
clean.
u
For integrated IDIS: check the light intensity.
ð
The displayed parameters must not blind the user or limit
the field of vision.
u
For integrated IDIS: check to make sure the total magnification
is correctly displayed. [
ð
The optical settings match the configuration on the surgical
microscope.
u
6.3.2 Function tests for device with fundus viewing system
Prior to using the fundus viewing system, reset the focus.
Action
u
With the fundus viewing system moved out of the beam path,
position the microscope in such a way that the index dot (A) of
the microscope's focal point is in the middle of the upper
triangle (B).
u
Select a medium magnification (i.e. 1.0).
u
Lower the microscope to the surgical field until the cornea of
the patient’s eye is in focus.
u
Tighten the [Downward travel limit] rotary knob.
u
With no patient present, check to make sure the suspension
arm cannot be lowered any further.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...