OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Operation
77
S7 floor stand
Balance setting
–
Check that the suspension arm has been properly balanced. When
fully balanced, the surgical microscope should remain stationary in all
positions within the work area and move neither upward nor down-
ward.
Limitation of downward travel
–
Check that the minimum working distance (height) from the surgical
field has been set.
Stand base
–
Check that two of the brake tabs have been pressed and prevent the
stand from starting to move by itself.
Accessories
–
Proper functioning of all other equipment (surgical microscope, coob-
servation equipment, video system, etc.) has been checked using the
relevant user's manuals.
Foot control panel
–
Check that the plug of the foot control panel has been connected.
–
Check that the power switch of suspension system has been switched
on.
•
Check that all functions assigned to the respective keys on the foot
control panel are working properly.
Warning!
If a function fails, you must not use this instrument for safety reasons. Cor-
rect the fault (see the "Troubleshooting table“) or contact our service dept.
Содержание OPMI 1FR XY
Страница 6: ...G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 22: ...22 Safety G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 38: ...38 Description G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 106: ...106 Technical data G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 112: ...112 Index G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007...
Страница 113: ...G 30 1416 en OPMI 1FR XY Surgical Microscope on S7 Floor Stand Issue 4 0 Printed on 14 11 2007 113...