Safety on the CMM
61211-1270302
MMZ M
2-17
Light barrier
Features of the light barrier
Different beam lengths
The light barrier is available with two different beam lengths:
–
0.5 m - 18 m
–
15 m - 70 m
Effectiveness
The light barrier is only effective in the AUTO operating mode. If the op-
erating mode switch is set to the MAN or 0 position, the light barrier will
not be effective.
Operating mode indica-
tion
The signal tower has three lamps used to indicate the operational states:
Color
Meaning
green
The light barrier is active and not interrupted. The
CMM moves at the speed defined in the measurement
plan.
white
The light barrier is not active. The MAN or 0 operating
mode has been selected.
Blue
EMERGENCY STOP has been triggered.
Functions of the light barrier
The light barrier may have two functions:
EMERGENCY STOP
and
REDUC-
TION
. The required light barrier function must be set by a ZEISS service
engineer during the start-up of the light barrier. It is
not
possible to
switch between the two functions.
Option 1:
EMERGENCY STOP function
An EMERGENCY STOP is activated when the light barrier is interrupted.
The drives are switched off and all travel movements are stopped. The
current measurement plan is interrupted and may need to be restarted
later if necessary.
Restarting
As soon as the light beam is no longer interrupted, the light barrier is ac-
tive again.
–
Switch on the drives.
Option 2:
REDUCTION function
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...