Special measures
61211-1270302
MMZ M
8-7
–
Reallocate the holder to the stylus system and choose the intermedi-
ate positions so that changes can be carried out without collision.
CMM in end positions
A limit switch is activated when the CMM moves to the end position.
The drives are switched off. In order to move the CMM again, the CMM
must be cleared by instructed personnel.
Clearing the CMM
WARNING
Injuries can be caused when safety devices are disabled.
Crushing and severing body parts.
ü
Safety devices have been disabled.
Clearing from the end position may only be performed by instructed
personnel.
Damage to the CMM by manual movements.
The CMM can only be moved out of the end position by means of the
joysticks. Never try to clear the CMM by sliding, pushing, or pulling
CMM components.
Move the CMM out of the end position by means of the joysticks on
the control console.
1
Disable the safety devices.
2
Switch on the drives.
Check in which direction the CMM must be moved before operating
the joysticks. The aim is to move the probe back into the measuring
volume. You must not move the CMM away from the measuring vol-
ume.
3
Clear the CMM using the joysticks.
4
Reactivate the safety devices.
NOTE
After clearing, a homing run has to be carried out.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...