Preparation for measuring operation
7-12
61211-1270302
MMZ M
–
Perform a reference measurement; use a master stylus.
–
Select the stylus to be qualified.
NOTE
Usually, the qualification process is carried out semi-automatically.
Please refer to the operating instructions for the measuring software.
Preparation
What is required for qualification?
The following is required for qualification:
–
Reference sphere
Usually, a high-precision ceramic sphere is used.
–
Master stylus
The master stylus must be used for the reference measurement.
➤
–
Styli required for workpiece measurement.
Information about the master stylus:
Red dot
The master stylus has different features with regard to length and diam-
eter of the stylus tip. The features depend on the probe used. Normally,
the master stylus is marked with a red dot.
The properties of master styli for the various probes are listed as follows:
Probe
Master stylus
Threads
Length
Diameter of the stylus tip
[mm]
[mm]
[mm]
VAST gold
M5
64
8
VAST XT gold
M5
64
8
VAST XTR gold
M5
64
8
VAST XXT TL1
M3
30
5
VAST XXT TL3
M3
30
5
VAST XXT TL4
M3
125
5
Requirements for proper qualification
Before starting qualification, make sure that the following prerequisites
are met:
–
The stylus is stable.
–
All stylus system components are screwed together tightly.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...