Components and function
61211-8140502
METROTOM 1500 G3 1.4
3-7
Gray scale value
The charge is converted into a voltage signal which corresponds to a
gray scale value on the monitor of the user computer. The gray scale
value can be set, among other things, by means of the detector gain
and the integration time.
Positioning system
The positioning system consists of several components: X guide beam, Y
slide, Z support, and a rotary table. There are four travel directions: X, Y,
Z, and R (rotary table axis).
Positioning system, travel movements
1 Coordinate system; origin near the X-ray tube
2 Y slide
3 Travel movement in X direction
4 Bellows cover on X guide beam
5 Travel movement in Z direction
6 Travel movement in Y direction
7 Z support
8 Rotary table axis