background image

 
Mechanical stage S mot. CAN 

INTRODUCTION 

Measuring stage S mot. CAN 

Notes on instrument safety 

Carl Zeiss 

 
 

 

08/2012 

435465-9000-701 

1.2

 

Notes on instrument safety 

 

The mechanical stage  /  measuring stage S mot. CAN may only be used for the applications 

described in this operating  manual. The manufacturer cannot assume any liability for other 

applications of the instrument, including those of individual modules or single parts.  

 

Modifications and repairs to this instrument and any devices operated in combination with the 

microscope must be carried out by our service department or by authorized personnel only. The 
manufacturer accepts no liability for damage caused by unauthorized access to the interior of 

the instrument. Failure to comply with this shall also render any warranty claims invalid. 

 

 

The device may only be operated by instructed personnel. The instruments may only be 

operated by trained personnel who are aware of the possible dangers involved with microscopy 

and the application concerned. The microscope is a high-precision instrument that can be 
impaired in its performance or destroyed when handled improperly. 

 

The mains plug of instruments with a metal housing (e.g. EMS 3) may only be connected to a 
socket with earth contact. The protective capacity must not be rendered ineffective by using an 

extension cable without a ground wire. 

 

If it is determined that protective measures are no longer effective, the instrument must be 
switched off and secured against inadvertent operation. Please contact a Zeiss service agency or 

the Carl Zeiss Microscopy Service to repair the instrument. 

 

Dust and dirt may impair the instrument’s performance. The instrument must be effectively 

protected from such influences and covered with the dust cover when not in use. Always check 
whether the instrument is switched off before you cover it. Broad temperature fluctuations, 

direct sunlight and vibrations should be avoided. 

 

 

The instruments are not equipped with special devices to protect against corrosive, potentially 
infectious, toxic, radioactive or other substances that may be hazardous to health. All legal 

regulations must be observed when handling such substances, particularly the relevant national 
accident prevention regulations. 

 

 

Defective instruments are not to be disposed of as ordinary domestic waste. They should be 
disposed of in accordance with the relevant regulations. 

 

Samples should also be disposed of in compliance with the relevant legal requirements and 

internal operating procedures. 

 

 

Before setting up the instrument, check whether it is suitable for the available line voltage. 

 

 

Always disconnect the instrument from the power outlet before opening it and changing the 

fuses. 
Only the specified fuses may be used.  

 

The use of makeshift fuses and short-circuiting of the fuse holders are not permitted. 

 

Controller EMS 3 supplies voltage to the mechanical / measuring stage S 150 x 100 mot. CAN. 

The EMS 3 permits mains voltages in the 100 to 240 V ±10% 50/60 Hz range without 

additional change in the voltage. 

 

Operation of the instrument in explosive environments is not permissible. 

Содержание Measuring stage S mot.

Страница 1: ...Operating Manual Mechanical Measuring stage S mot for Axio Zoom and SteREO Discovery...

Страница 2: ......

Страница 3: ...d a high quality product from Zeiss Before using the instrument the first time please read this operating manual in order to maintain the high quality of the instrument and ensure reliable work with i...

Страница 4: ...anual is not subject to updating or revision Unless explicitly authorized dissemination or duplication of this document and commercial exploitation or communication of its contents are not permitted P...

Страница 5: ...1 General information 14 3 2 Mounting the mechanical measuring stage S mot CAN 14 3 2 1 Mounting the mechanical measuring stage S mot CAN on base 300 for epi illumination 15 3 2 2 Mounting the mechan...

Страница 6: ...magnetic Compatibility and 2006 42 EC Machinery Directive with the appropriate controller EMS 3 The operating manual contains all information and warnings the operator must comply with The following w...

Страница 7: ...itched off and secured against inadvertent operation Please contact a Zeiss service agency or the Carl Zeiss Microscopy Service to repair the instrument Dust and dirt may impair the instrument s perfo...

Страница 8: ...P 3 is at the zero position and not swiveled out and that the foot pedal is clear Automatic travel of the motorized tables can be interrupted by pressing the STOP button on the focusing drive or the S...

Страница 9: ...wear wearing parts in particular and improper use The instrument manufacturer shall not be liable for damage caused by faulty operation negligence or any other tampering with the instrument particular...

Страница 10: ...es When using the mechanical measuring stage S mot CAN the separate operating manuals for the Axio Zoom V16 or SteREO Discovery and insofar as they are used the Transillumination top 450 mot SYCOP 3 s...

Страница 11: ...010 000 EMS 3 Required for operating the stage USB 2 0 cable to workstation with software Country specific power cable SYCOP 3 system control panel 435611 9010 000 SYCOP 3 for stage control Cable MDR...

Страница 12: ...Objective E Nosepiece F Intermediate tube Fluar Illuminator Z mot G Camera adapter with microscope camera AxioCam H Reflected light illumination slit ring illuminator with cold light source CL 9000 L...

Страница 13: ...kaging Permissible ambient temperature 5 to 40 C Permissible relative humidity max 75 at 35 C non condensing Transport in packaging Permissible ambient temperature 40 to 70 C Operation Permissible amb...

Страница 14: ...l shipping braces adhesive tape or similar Keep the original packaging for storage or for returning the instrument to the manufacturer or dispose of it properly 3 2 Mounting the mechanical measuring s...

Страница 15: ...o this push the plate Fig 4 3 into the stage Fig 4 1 with the left front corner slightly inclined allow it to slide under slight pressure against the springs Fig 4 4 until it is completely inside the...

Страница 16: ...into the stand base 450 and secure with the three captive screws Fig 5 6 Detach the insert plate Fig 5 3 from the stand base 450 Fig 5 5 Set the mechanical measuring stage Fig 6 3 on the insert plate...

Страница 17: ...w it hand tight with the 4 captive screws Set the mechanical measuring stage Fig 7 5 on the transillumination base 300 stage drive pointing to the left align to the rear edge and secure with the three...

Страница 18: ...4 Remove the three short hexagon socket screws Fig 8 5 on the insert plate and keep for possible later use Remove assembly aid d 120 mm Fig 8 2 again Set the mechanical measuring stage Fig 8 7 on the...

Страница 19: ...tension cable HD15 Fig 9 3 via the upper connecting socket Fig 9 1 Connect SOCP 3 Fig 9 5 to EMS 3 In addition to or as an alternative to SYCOP 3 the mobile coaxial drive XY CAN can be connected using...

Страница 20: ...sed while moving the joystick accordingly If SYCOP 3 is not available on the microscope system the mobile coaxial drive XY CAN can be optionally used for external control of the stage see Section 4 2...

Страница 21: ...p Mode Stage Calibrate Calibration is necessary for use of the SYCOP 3 memory function It is recommended that this is carried out immediately upon prompting The settings in the Memory menu are thus av...

Страница 22: ...positioned left and right Fig 11 2 or 3 Saving the current XY position by holding down the respective button Going to the saved XY position by briefly pressing the respective button The XY positions a...

Отзывы: