SET-UP
Mechanical stage S mot. CAN
Carl Zeiss
Mounting the mechanical / …
Measuring stage S mot. CAN
18
435465-9000-701
08/2012
3.2.4
Mounting the mechanical /
measuring stage S mot. CAN on the
transillumination top DL 450 mot.
The procedure for mounting trans-
illumination top DL 450 mot. on base
plate 450 is described in the separate
operating manual of the trans-
illumination top.
•
Remove the glass plate (Fig. 8/
1
) from insert
plate (Fig. 8/
3
) by pressing and tilting the rear
edge.
•
Insert assembly aid d = 120 mm (Fig. 8/
2
) to
prevent the screws and small parts from being
lost in the transillumination top (Fig. 8/
4
).
•
Remove the three short hexagon socket screws
(Fig. 8/
5
) on the insert plate and keep for
possible later use.
•
Remove assembly aid d = 120 mm (Fig. 8/
2
)
again.
•
Set the mechanical / measuring stage (Fig. 8/
7
)
on the insert plate (stage drive pointing to the
left), align to the rear edge of the
transillumination top (Fig. 8/
4
) and secure on
the transillumination top with the three captive
screws (Fig. 8/
6
) of the stage.
•
Insert the glass plate (Fig. 8/
8
) into the stage.
The procedure for inserting the plate
into and removing from the stage is
described on page 15.
The red dots on the table (Fig. 8/
7
) and
glass plate (Fig. 8/
8
) are to ensure
reproducibility of the alignment of the
metal plate and stage if the former
needs to be detached.
Fig. 8
Mounting the
mechanical /
measuring stage mot.
on
transillumination top DL 450 mot.