Left Tool Area and Hardware Control Tools
LSM 880
322
000000-2071-464
10/2014 V_01
The
Sampling
mode is a meant to improve the image display as it equalizes the differences between the
very high and the very low intensities in the image. However, it is not suitable for quantitative analysis as
the level steps become non-linear.
The image re-calculation works already with a single image (
Frame
= 1) and is done in the following way
taking
Intensity
3.
Image x 1; cut off of grey levels above the dynamic range; Result divided by 1
Image x 2; cut off of grey levels above the dynamic range; Result divided by 2
Image x 3; cut off of grey levels above the dynamic range; Result divided by 3
The three resulting images are summed up and again divided by 3. This procedure is adapted accordingly
if a higher
Intensity
number is set. If two images are
acquired
(
Frame
= 2) and Intensity is still at 3, the
second image is rated lower and is multiplied and divided in three steps with the value 4,5 and 6; The
then 6 resulting images are summed up and divided by 6.
The image acquisition parameters are not changed for the acquisition of the two (or more)
Frames
.
Fig. 426
HDR image display (left side without, right with HDR)