LSM 510
ANNEX
LSM 510 META
Monitor Diode
Carl Zeiss
03/06
B 45-0021 e
8-21
The following image is an example of the reduction of correlated noise. The low frequency noise has
been generated artificially.
The image in the upper left corner shows the PMT image plus noise, the image beneath this (upper right
corner) shows the signal of the diode expanded to 512 x 512 pixels (noise without object information).
The two images below show the ratio of the PMT and diode signal (left) and the sum of all signals (right).
The sum-image does not contain any information and can therefore be neglected.
To get a ratio image like the one shown here, Detector Gain, Amplifier Gain, Amplifier Offset of the PMT
channel, Gain and Offset of the diode channel, Gain and Offset of the ratio channel must be set in the
correct way.
Each of the parameters summarized effects either the amplification of the ratio image, or the contrast of
the ratio image, or the quality of the noise reduction.
Fig. 8-8
Image Display window
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......