8 ANNEX
8.1
Beam Path Configuration Guide for Fluorescence Microscopy with the LSM 5
8.1.1
Optical elements in the Configuration Control window
All wavelength values given in Nanometer [nm].
8.1.1.1
Main Dichroic Beam Splitter (HFT)
•
A
HFT XXX[/YYY/...]
deflects the indicated laser lines onto the specimen and allows the emitted
fluorescent light to pass through.
Example: HFT 458/514, HFT UV/488/543/633 (deflects also UV excitation light)
•
A
HFT KP XXX
(KP = Short Pass) is a special type of a main dichroic used for IR multiphoton
excitation. The
HFT KP 650
, deflects laser light with a wavelength longer as 650 nm onto the
specimen and allows fluorescent emission light in the visible range below 650 nm to pass through.
Example: HFT KP 650
•
A
HFT KP XXX_YYY
is a combination of a HFT YYY and HFT KP XXX used for simultaneous IR multi
photon and single photon excitation.
Example: HFT KP 700_488.
8.1.1.2
Secondary Dichroic Beam Splitter (NFT)
•
The
NFT XXX
is used to split the emitted light which will be guided into separate channels. Light with
shorter wavelength than XXX nm is deflected, light with longer wavelength passes the NFT. A cascade
of NFTs allows to distribute the emission light to more than two channels/detectors.
•
The
NFT KP YYY
treats the emission light the other way round: it transmits light shorter than YYY nm
and deflects above YYY nm.
8.1.1.3
Emission Filters (EF)
•
A
LP XXX
(Long Pass) transmits emission light with wavelengths longer than the indicated threshold
value XXX.
•
A
KP XXX
(Short Pass)
transmits emission light with wavelengths shorter than the indicated threshold
value XXX.
•
A
BP XXX-YYY
(Band Pass) transmits emission light within the indicated wavelength band.
•
A
BP XXX/BB
has a transmission band for emission light with a center wavelength of XXX nm and a
width of BB nm.
•
The
BG 39
(Blue Green glass) blocks infrared excitation light by absorption.
•
BP ... IR
(Band Pass – Infra Red)
is a band pass suitable for detection of IR excited dyes. It blocks the IR
light.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......