4.13.24.8 3D Measurement Functions
(1)
Volume measurement mode (Flood function)
•
Use the 3D button to select the required 3D display of the stack.
•
Click on the
Volume
button in the
Measure
button bar.
−
The volume parameters are calculated and displayed below the image.
−
The
Copy
button is displayed below the right-hand side of the image. This button permits the
volume values to be copied to the clipboard and imported to other programs (e.g.: MS Word or MS
Excel) via the
Paste
function.
•
Setting the
Fill Level
slider enables you to change the height level of the topography. The portion of
the topography lying below the set height level is filled with "water" (blue color) and the volume
parameters are calculated online only for the projecting part of the topography.
To use the
Fill Level
function, load the
Profiles
3D display mode containing the
Glowscale
palette, or activate
No Palette
to obtain optimum display.
•
If the
Diagram
function
Curve of tp
is also activated, a red marker line shows the position of the
height level in the percentage of contact area curve.
Fig. 4-306
Image Display window, Topography display: 3D - Volume
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......