OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Maintain Menu
LSM 510 META
DuoScan
4-238
B 45-0021 e
03/06
4.10.3.2 Pinhole
Panel
No further software function can be activated and executed during pinhole adjustment.
Pinhole / Description
field:
Selection of pinholes (PH1 to PH4) to be adjusted via the
Pinhole
selection
box, display of the relevant channel in the
Description
field.
Diameter; Pos. X; Y; Z
slider:
Setting of diameter, X-, Y- and Z-position of the pinhole in relation to the
beam path (Z-position can be set only for PH1) using the slider or arrow
buttons, status display for setting procedure: green for ready and red for
busy.
Store current Position
button:
Storage of the current pinhole setting.
Move to stored Position
button:
Pinhole setting is reset to the position last stored.
Adjust Automatically
button:
Automatic pinhole adjustment.
Fast Adjust mode
check
box:
If this check box is activated, the pinhole adjustment is only performed in a
limited area. Used for readjustment.
Fig. 4-238
Pinhole panel
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......