LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Options Menu
Carl Zeiss
03/06
B 45-0021 e
4-221
4.9.5.9
Print Status Display
This tab permits selection of which information is
displayed in print preview.
On activation of the
Print Status Information
check box, the status information will be printed.
4.9.5.10
Recording / Reuse
The parameters to be taken into consideration for
the use or load of a recording configuration are set
in this tab.
As an option, you can also determine whether the
objective setting shall be taken over when the
Reuse
function is used.
4.9.5.11 Time
Series
In the
Time series
tab, you can determine
whether the time for the recording of a time series
is set as
Cycle Delay
or as
Time Interval
.
Cyle Delay
is the interval between the end of one
scan process and the beginning of the next.
Time Interval
is the interval between the
beginning of one scan process and the beginning
of the next.
You can select the unit for
Mean of ROIs
diagrams.
Fig. 4-214
Print Status Display tab
Fig. 4-215
Recording / Reuse tab
Fig. 4-216
Timeseries tab
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......