LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Acquire Menu
Carl Zeiss
03/06
B 45-0021 e
4-87
Mark First/Last
tab
The determination of the optimum stack size is
performed here via focusing during a continuous
scan.
•
Click on the
XYcont
button.
−
A continuous XY-scan of the set focus
position will be performed.
−
If you have reduced the scan speed or have
set image averaging, you should use the fast
scanning mode to find the lowest and
highest points of focus. These settings are
made under
Mode
in the
Scan Control
menu, or directly via the
FAST XY
button.
•
Use the manual focusing drive or the
Stage and Focus Control
window (see
Stage
, page 4-122) to
focus on the upper position of the specimen area where the Z Stack is to start.
•
Click on the
Mark First
button to set the upper position of the Z Stack.
•
Then focus on the lower specimen area where the recording of the Z Stack is to end.
•
Click on the
Mark Last
button to set this lower position.
•
Keep Slice
enables the
Num Slices
slider which allows to vary the number of slices. The limits of the
Z Stack remain constant, the interval is matched accordingly.
•
Keep Interval
enables the
Interval [µm]
slider which allows to vary the interval between the slices.
The number of slices is matched accordingly, the limit of the Z Stack is adjusted.
•
Click on the
Start
button to start the recording of the Z Stack.
In case the upper and lower limits of the stack have been switched round, automatic matching will be
performed by the software, since the stage of the Axio Imager.Z1 always moves from bottom to top and
the nosepiece of the Axiovert 200 M always moves from top to bottom.
Setting
via
Range
is not possible via the
Mark First/Last
function, i.e. the lines cannot be
shifted.
The
Fast Z Line
functions is not available in frame mode.
When you change from
Mark First/Last
to
Z Sectioning
or vice versa, the values are updated
in the
Z Sectioning
tab.
Fig. 4-67
Mark First/Last tab
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......