OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-56
B 45-0021 e
03/06
4.5.3.5
Settings for Single Track in the Channel Mode
The settings of the beam path for the scanning procedure with regard to the main dichroic beam splitter
(HFT), secondary dichroic beam splitter (NFT), emission filters (EM) to be used and the assignment of
channels, excitation wavelengths and laser intensities are performed in the
Beam Path and Channel
Assignment
panel.
The setting can be performed manually or by using existing track configurations.
•
Click on the
Single Track
button, unless it has already been activated.
−
The
Configuration Control
window for single tracking is displayed.
•
Click on the
Descanned
button, unless it has already been activated.
(1)
Beam Path and Channel Assignment panel
The
Beam Path and Channel Assignment
panel displays the selected track configuration which is used
for the scan procedure.
You can change the settings of this panel using the following function elements.
Activation / deactivation of the excitation wavelengths (check box) and setting of
excitation intensities (slider). Open the
Laser Control
window via the
Laser
button.
Selection of the main dichroic beam splitter (HFT) or secondary dichroic beam splitter
(NFT) position through selection from the relevant list box.
Selection of an emission filter through selection from the relevant list box.
Activation / deactivation of the selected channel (Ch 1-4, monitor diode ChM, META
detectors ChS1-8, transmission ChD) for the scanning procedure by assigning an
existing color icon or defining a new one. Deactivation of the channel via deactivation
of the check box.
The
Fiber Out
port of the LSM 510 permits connection to another detector.
For the configuration of the beam path, please refer to the application-specific configurations depending
on the used dyes and markers and the existing instrument configuration (e.g.: module LSM 510 - Config.
26) listed in the annex.
The assignment of the numbered emission filters (1-4), NFT secondary dichroic beam splitters (1-3) and
HFT main dichroic beam splitters in the
Beam Path and Channel Assignment
panel is shown in the
Configuration Control
window (Fig. 4-39). The numbers of the emission filters correspond to those of
the channels lying behind (PMT photomultipliers).
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......