Lightsheet Z.1
Maintenance and Cleaning
Carl Zeiss
02/2013
000000-1790-528
3
1
Maintenance and Cleaning
The maintenance to be carried out by the customer is limited to cleaning the painted surfaces, replacing
glass windows and rubber seals on the sample chamber, cleaning and disinfecting the system cavity and
sample chamber, temperature sensors and detection optics unit, replenishment of coolants for the
incubation device and liquid cooling of the detection module as necessary, as well as the replacement of
CO
2
sterile filters for the incubation device.
a) To clean painted surfaces proceed as follows:
•
Switch the device off completely and pull the mains plug.
•
Make sure that no cleaning fluid is allowed to enter the system.
•
Wipe the painted surfaces with a clean cloth moistened with water to which a small amount of
cleaning agent has been added. Do not use a solvent. Dry off with a lint-free cloth.
b) To clean and disinfect the system cavity proceed as follows:
•
Switch the device off completely and pull the mains plug.
•
Make sure that no cleaning fluid is allowed to enter the system.
•
Take care not to touch the front lens of the illumination and detection optics units. Remove these
optics as necessary.
•
Remove the sample chamber for cleaning.
•
Cleaning the system cavity:
Wipe the inner surfaces of the system cavity with a clean cloth moistened with water to which a
small amount of cleaning agent has been added. Do not use a solvent. Dry off with a lint-free
cloth.
•
Disinfecting the system cavity:
Carefully wipe the inner surfaces of the sample chamber with a lint-free cloth soaked in 70
%
ethanol or 70
% isopropyl. Allow the solution to react for a few seconds. Repeat the procedure
once or twice as necessary. Carefully wipe off the inner surfaces with a lint-free cloth soaked in
distilled water.
Содержание Lightsheet Z.1
Страница 1: ...Lightsheet Z 1 Operating Manual February 2013 ZEN 2012 black edition ...
Страница 4: ......
Страница 170: ......
Страница 427: ...Lightsheet Z 1 Overview ...