Carl Zeiss
Configuration of the Light Path
Lightsheet Z.1
24
000000-1790-528
02/2013
You can change the imaged area using the
Zoom
function. This zoom utilizes hardware zoom
optics within the system.
Within the
Zoom range
from 0.36 to 0.7, the image quality towards the edges can decrease
(
Zoom
numbers are displayed in
red
; range is indicated with a
red
line below the slider).
Be aware that when you change the Zoom during an acquisition session, the light sheet alignment
(Lightsheet
Auto-Adjust
in the
Channel tool
window) needs to be re-done.
The
Single Side
illumination is displayed by default from the
Left
direction irrespective of the
configuration. This setting corresponds to the standard configuration where the light sheet only
illuminates the sample from one side.
If Single Side is chosen for Illumination, the direction,
Left
or
Right
, must be marked. When Dual
Side illumination is chosen the sample is illuminated by the light sheet from the left and by the light
sheet from the right sequentially, producing two images for each plane.