1 About these Installation Requirements | 1.1 Further Applicable Documents

ZEISS

1 About these Installation Requirements

This document describes the location requirements for the installation of the Lightsheet 7, here-
inafter referred to as Microscope System.

It helps to determine a suitable location within facilities and to prepare the location before deliv-
ery of the Microscope System. The suitability of a location depends on various factors such as:

§

Room size

§

Floor stability

§

Environmental conditions

§

Availability of supplies

The room size should be able to accommodate the Microscope System as well as potential acces-
sories and 3rd party systems that may be attached to the Microscope System. For servicing the
equipment, adequate access space is also required. The selected location should conform with the
correct environmental conditions to ensure the safe and ideal operation of the Microscope System
and its application whilst still ensuring the safety and comfort of the operators. Environmental
conditions are often influenced by surrounding installed systems or activities. We can distinguish
between:

§

Vibrations

§

Acoustic noise

§

Room temperature

§

Air quality

The customer is responsible for ensuring that the conditions specified in this document are met
and that the required equipment is available before delivery and installation. A ZEISS service repre-
sentative will gladly assist you in examining the installation conditions and suitable measures.

1.1 Further Applicable Documents

Please take also note of the following documents:

Downloads

Brochures

, Safety data sheets, ISO certificates, CSA certificates and EU declarations of conformity

as well as software and firmware updates can be found at 

https://www.zeiss.com/microscopy/int/

downloads.html

 or ask your ZEISS Sales & Service Partner.

Local and National

health and safety

regulations

Local and national health and safety regulations should be adhered to for the site and the use of
the Microscope System. Please consult with your ZEISS Sales & Service Partner in the event of
these regulations conflicting with the installation requirements of the Microscope System.

System and 3rd

Party Components,

Accessories

The Microscope System can be configured in many ways. Information about the individual com-
ponents, enhancements and accessories can be obtained from your ZEISS Sales & Service Partner.
Also refer to the 3rd Party documentation of the manufacturer.

4

Installation Requirements ZEISS Lightsheet 7 | en-US | Rev. 1 | 349500-8186-000

Содержание Lightsheet 7

Страница 1: ...Installation Requirements ZEISS Lightsheet 7 ...

Страница 2: ...es for archiving purposes shall remain unaffected thereby Any viola tions may be prosecuted as copyright infringements The use of general descriptive names registered names trademarks etc in this document does not imply that such names are exempt from the relevant intellectual property laws and regulations and therefore free for general use This shall also apply if this is not specifically referre...

Страница 3: ...gory 10 2 3 2 UPS Requirements 11 2 4 Network Connectivity 12 2 4 1 ZEISS Predictive Service 12 2 5 Safety and Radiation Information 13 2 5 1 Laser Safety 13 2 5 2 Biosafety Level of the Laboratory 14 2 6 Gas Supplies 14 2 6 1 Compressed Air 15 2 6 2 CO and other Gas Supplies 15 2 7 Cooling System 16 2 8 Transport and Storage 16 2 8 1 Weight and Sizes of the Transported Goods 16 2 8 2 Weight and S...

Страница 4: ...emperature Air quality The customer is responsible for ensuring that the conditions specified in this document are met and that the required equipment is available before delivery and installation A ZEISS service repre sentative will gladly assist you in examining the installation conditions and suitable measures 1 1 Further Applicable Documents Please take also note of the following documents Dow...

Страница 5: ...0 dB Z class 200 to 500 Hz less than 55 dB Z class 2 1 2 Air Conditioning and Quality The following nominal power usage for the Microscope System and possible accessories is shown in Watts To maintain a more constant room temperature the size of the air conditioning should be well matched to the size of the room and the heat generated The stable air conditioning of the room is critical to achievin...

Страница 6: ...llution occurs Occasionally temporary conductivity caused by condensation is to be expected Example A circuit used in an of fice environment The circuitry inside a computer would fall in this cate gory 3 Conductive pollution occurs or dry nonconductive pollution occurs that becomes conductive due to condensation that is to be expected Example Circuitry that is exposed to outside air but will not b...

Страница 7: ...tem must only be operated in closed rooms It is recommended to install the Microscope System in a dimmed room where artificial illumination sunlight or other light sources cannot interfere with image acquisition The Microscope System should not be installed near radi ators or windows with direct sunlight Compliance with the installation requirements of the Mi croscope System and the availability o...

Страница 8: ...asurements in mm For possible changes in pref erential layout please discuss with your ZEISS representative The placing of the user table can vary depending on available space and cable lengths Data cable max 30 m 500 ca 1600 750 800 500 900 1200 approx 2300 1200 5 4 9 8 6 7 2 1 3 850 incl corner pieces 1000 incl corner pieces 500 Fig 1 Space requirement for Lightsheet 7 with a storage and data an...

Страница 9: ...E 220 240 V AC 10 100 125 V AC 10 Nominal frequency 50 60 Hz Main Power Plug Local mains plug will be supplied Power consumption 220 240 V 800 VA 100 125 V 750 VA Storage PC Max 400 VA Max current 3 5 A at 220 V 8 A at 100 V Additional building PE The system must be connected to a building earth point at all times Additional mains power 3 x Standard local 230 or 120 V mains sock ets Required for t...

Страница 10: ...ation of motors generators or the switching of reactive circuit components Random transients on the other hand are often caused by lightning and electro static discharge Ensure that the possibility of transient voltages on the mains line are minimal to prevent damage to the system Max transient voltage in V Nominal Voltage VAC Category I Category II Category III 50 330 500 800 100 500 800 1500 150...

Страница 11: ...ements If the site requires a UPS Uninterruptable Power Supply please ensure the KVA rating of the UPS meets the ratings of the Microscope System and all third party items that may also be connected e g water chiller air compressor sample prep equipment EDS or WDS etc This will ensure that the complete system will continue to function during a blackout or brownout long enough to allow the user to ...

Страница 12: ...d for condition monitoring by systematic retrieval of relevant instru ment health information and its server based processing The goal is to detect deviations in the in strument performance before they impair the user s result or even result in unplanned downtime For details ask the ZEISS service representative about recommendations for data security policy on ZEISS Systems For predictive service ...

Страница 13: ...safety regula tions The following types of lasers are currently intended for use Using other lasers than indicated be low requires prior coordination with ZEISS The performance values of the directly integrated mul tiphoton lasers correspond to the rated laser power Laser Wavelength nm Class Typical power ex fiber mW solid state laser SSL 405 3B 20 solid state laser SSL 405 3B 50 solid state laser...

Страница 14: ...er some do exist in a small number of places in the US and around the world As the highest level of biological safety a BSL 4 lab consists of work with highly dangerous and exotic mi crobes Infections caused by these types of microbes are frequently fatal since there are no available treatments or vaccines Two exam ples of such microbes include Ebola and Marburg viruses Tab 3 Biosafety Level 2 6 G...

Страница 15: ...user if the air supplied is from the building facilities or from gas bottles Due to acoustic noise and vibrations where possible the compressor should be installed in a sep arate room close to the workstation Make sure there is enough room for air circulation to the sides and at least 100 cm to the back of the compressor Refer to the manufacturer s documenta tion for further details 2 6 2 CO and o...

Страница 16: ...he weight information on the package and on the shipping document Where possible the original packaging must be used for shipping or transport Fork lift and a hand pallet truck For on site transport and unloading a fork lift and a hand pallet truck are necessary Ensure all hallways and corners are wide enough to be passed with the hand pallet truck Please check the location requirements for door a...

Страница 17: ...ves some indication on the approximate weight and sizes of the unpacked items Component Length mm Width mm Height mm Weight kg System table small incl corner pieces 900 1000 750 850 810 870 130 Main system module Light sheet 7 800 450 500 75 LB Rack Lightsheet 700 600 550 80 Cooling unit INNOVATEK 3156 01 500 130 190 7 5 Info The tables for the computers are provided by the customer ZEISS recommen...

Страница 18: ... international guidelines and standards the 21 CFR 1040 10 Performance Standards for light emitting products laser products applies for the USA The following EMC user notice is for Korea only 기종별 사용자안내문 A급기기 업무용방송통신기자재 이기기는업무용 A급 전자파적합기기로서 판매자또는사용자는이점을주의하시기바라 며 가정외의지역에서사용하는것을목적으 로합니다 European and International Directives Standards For more information on ISO and CSA certifi cates CE Declarations o...

Страница 19: ...h requirements are yet to be completed ZEISS can be requested to assist with the checking of the environmental conditions Environmental conditions in and around a system can change with time due to various changes to the building new equipment being installed or other reasons Should the environmental condi tions change due to changes to the building or other potential external influence after a si...

Страница 20: ...Radiation Information X Ray Laser Safety Biosafety Level of the Laboratory Cooling Water Supply Cooling liquid and or chemicals Exhaust Line Gas Supplies Compressed Air Nitrogen Additional Gas Transport and Storage Place date Name Signature The above signature conveys that the Installation Requirements have been read and understood The customer further understands that some of these requirements a...

Страница 21: ......

Страница 22: ...Zeiss Promenade 10 07745 Jena Germany phone 49 1803 33 63 34 fax 49 3641 64 3439 info microscopy de zeiss com www zeiss com microscopy Installation Requirements ZEISS Lightsheet 7 en US Rev 1 Modifications reserved Printed in Germany ...

Отзывы: