5 Commissioning and First Operating Steps | 5.7 Finding Appropriate Detector Settings
ZEISS
à
In the upper section, the
BSD Control
dialog
displays the four quadrants/outputs.
5. Click a quadrant symbol to toggle its status between on (white), inverted (black), and off
(gray).
6. To confirm the settings, click
Apply
.
7. To choose the respective mode, click
BSD: COMPO
or
BSD: TOPO
.
The default setting for BSD is
BSD: COMPO
. All four segments are set to
on
and an image
is obtained, which is high in atomic number contrast.
8. If you want to change the default setting to
BSD: TOPO
, click
BSD: Set TOPO
.
9. From the
BSD Gain
drop-down list, select the appropriate BSD Gain:
Low
,
Medium
,
High
, or
Very High
.
The BSD Gain depends on the signal strength.
5.7.7 Setting up the AsB4 Detector
The AsB4 detector is optionally available. When ordered this detector replaces the AsB detector.
The AsB4 detector enables four parallel outputs for simultaneous compositional and topographi-
cal imaging. Each of the four channels can be selected separately.
The following settings are recommended for the AsB4 detector:
Detector Settings
1.5 kV – 30 kV
3–10 mm
§
Use compositional mode for obtaining an
image that is high in atomic number con-
trast.
§
Use topography mode for showing sur-
face details.
§
Use individual settings for channeling
contrast.
Procedure
1. In the GeminiSEM Control panel, select the Imaging tab.
2. From the
Signal A
drop-down list, select
AsB4 Ch1
.
3. From the
Tool Bar
, select
Quad Mode
.
96
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000