Operation
13
626109-9160-000 FixAssist XTR Operating manual V1.0
4.2
Configuring the angle around the Z-axis
1.
2.
Release the four screws (
1
)
slightly with a torque
screwdriver until
the probe system can rotate.
Turn the probe system
towards the prism.
1
1
Fig. 8:
3.
4.
Release the wing nuts (
2
) of
the angular scale and adjust
the desired angle by rotating
the angular scale.
In this process, think about
in which direction you want
your probe system to turn.
Re-apply the wing nuts. The
angle of the probe plate is
set.
2
Fig. 9:
5.
6.
7.
Set the prism vertically.
To do this, release the fixing
screws (
3
) and adjust the
prism angle 0°.
Re-apply the fixing screws
again.
Release the yellow fixing
screw (
4
) of the horizontal
guide and push the prism
slide carefully toward the
probe until it is against the
prism.
3
4
Fig. 10: