7 Troubleshooting | 7.2 Column
ZEISS
4. Close the gun.
5. Reinstall the top cover.
à
The microscope is completely reassembled.
6. Energize the microscope, start the microscope and SmartSEM, refer to
7.2.2.5 Startup of a LaB₆ Filament
7.2.2.5.1 Starting up the LaB₆ Filament for the First Time
Before you can use the filament for SEM imaging, you need to evacuate the microscope and start
up the filament.
Prerequisite
ü
You have energized and started the microscope.
Procedure
1. In the
SEM Controls
panel, select the
Gun Vacuum
tab.
2. Evacuate the microscope via
Pump
.
3. Wait until the messages
Vac Status = Ready
and
EHT Vac ready = Yes
are displayed.
INFO:
It may take some time until the vacuum is established and the messages are dis-
played.
4. In the
Gun
tab of the
SEM Controls
panel, deactivate
Long Fil. Life
.
5. Switch on the electron beam.
6. Set the following values:
EHT = 10.00 kV
Spot Size = 500
Fil I Target = 1950 mA
144
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006