5 First Operating Steps | 5.7 Working with Specific Specimen Types
5.7.3.3 Activating the Peltier Coolstage
You need to activate the stage and the EP option in the software and set the working pressure
before you can use the EP mode.
NOTICE
Rotating the Peltier coolstage
The Peltier coolstage can be damaged when stage rotation is used.
4
Only move the stage in X and Y direction.
Procedure
1. Switch on the Peltier coolstage power supply.
2. To evacuate the system, click
Pump
in the
Vacuum
tab of the
SEM Controls
panel.
3. Wait until
Vac Status = Ready
and
EHT Vac ready = Yes
are displayed.
This can take some time.
4. From the
Menu Bar
, select
Stage > Navigation
.
à
The
Stage Navigation
panel is displayed.
5. Click
Settings
.
à
The
Stage Navigation Settings
dialog is displayed.
6. In the
Stage Navigation Settings
dialog, click
Show Holder Gallery
.
à
The
Sample Holder Gallery
dialog is displayed.
7. In the
Sample Holder Gallery
dialog, click
CoolStage MK3
.
8. Activate the
Is Available
checkbox.
9. Close the
Sample Holder Gallery
dialog.
10. Close the
Stage Navigation Settings
dialog.
11. Select
Tools > Administrator
and log in with your username and password.
12. Select
Other
and activate the
Peltier Fitted
checkbox.