Lamp replacement and calibration - Corona process | Service manual |
EN_40_070_0010I
Sensor calibration
Lamp replacement and calibration Corona process
6
Corona process alignment
Caution
SENSITIVE COMPONENTS
Do not touch the white surface of the white standard with your fingers.
Protect the white standard from dust and impurities. Cover it up with
the delivered lid when not using it.
We recommend certifying the white standard once a year.
Place the white standard in the standard holder at position
P6
.
Fig. 35 Mounting the white standard
Align the
Corona process
with the adjustment screws
6
(see
“Fig. 37 Positioning the white standard” on page 37). Ensure that the light
beam is shining on the center of the white standard. Click
Continue
after
aligning the
Corona process
.
Fig. 36 Confirming the alignment
This procedure may take a few seconds.