NOTES ON INSTRUMENT SAFETY
Carl Zeiss
Safety Instructions
Cell Observer SD
8
423638-7044-001
M60-2-0021 e 01/2009
1.2.2
Installation Instructions
The Cell Observer SD system may only be installed or set-up by Carl Zeiss MicroImaging GmbH
Customer Service technicians or personnel specially authorized by the Manufacturer.
Carl Zeiss MicroImaging GmbH Customer Service will provide such services on a chargeable
basis.
Do not install the Cell Observer SD near heat sources, e.g. radiators or direct sunlight. To avoid
build-ups of heat, the ventilation slits at the microscope system must not be obstructed.
Do not operate the system, including standard-supplied accessories, in an explosive atmos-
phere, in the presence of such volatile anaesthetics or flammable solvents as alcohol, gasoline
or similar substances.
The main power plug may only be inserted into a socket with PE contact, and its built-in
protection device must not be rendered ineffective by using an extension cable without PE
conductor.
The Cell Observer SD components must be powered via two or three separate line power
cables, not via a common power distribution strip. This is necessary to prevent excessively
strong leakage currents (refer to section 3.1.2 on page 38).
The earth cable which is included in delivery (AWG10 green/yellow) must be used for
connecting the multi laser module to earth-ground. An earth-ground point must be available in
the installation room.
Check that your actual voltage levels correspond to the voltages that are indicated on the
nameplate of the laser module before you connect a line power cable.
For laser safety reasons, all TV outlets at the microscope must be connected to a camera or
sealed with a cap.
Maintenance or repair, including work for modification, removal or replacement of components
other than or in excess of the work scope described in this Operating Manual, as well as any
type of intervention into the inner equipment space may only be performed by technicians of
Carl Zeiss MicroImaging GmbH or persons expressly authorized to handle such work tasks on
behalf of Carl Zeiss MicroImaging GmbH. In particular, this applies to the microscope system,
the CSU-X1 spinning disc unit, the PC system and to cable connections.
Please note that the Cell Observer SD is an opto-electronic precision product. Improper
handling may easily affect its function or even damage the system. All claims for warranty will
be null and void in this case.