4 Installation | 4.12 Assembling the Reflected Light Light Source
4.12 Assembling the Reflected Light Light Source
4.12.1 Assembling the LED10 Light Source for Reflected Light Illumination
For assembling the LED10 light source for reflected light illumination, proceed in the same way as
for the
LED10 light source for transmitted light illumination [
4.12.2 Assembling the HAL 100 Light Source for Reflected Light Illumination
For assembling the HAL 100 light source for reflected light illumination, proceed in the same way
as
transmitted light illumination [
.
4.12.3 Assembling the Adjustment Tool of HBO 100
This procedure applies to the following stand type only:
§
Axioscope 5 Bio-TL/RL
1
4
3
2
Fig. 47: Assembling the adjustment tool
1
Cover cap
2
Screw (3x)
3
Adjustment tool
4
Cover cap with opening
Prerequisite
ü
The microscope is switched off.
ü
The HBO 100 light source is installed.
Procedure
1. Remove the cover cap
1
from the mounting aperture of the stand.
2. Insert the adjustment tool
3
.
3. Tighten the three included screws
2
.
4. Attach the cover cap with opening
4
to the mounting aperture. Make sure it locks.
5. Slide in the movable connecting piece of the adjustment tool.
Proceed in the reverse order for removal.