![Zeiss Axioscope 5 Скачать руководство пользователя страница 76](http://html2.mh-extra.com/html/zeiss/axioscope-5/axioscope-5_instruction-manual_3365198076.webp)
4 Installation | 4.11 Assembling the Transmitted Light Light Source
ZEISS
2
4
6
3
5
7
1
8
1
Fig. 46: Assembling the HAL 100 light source
1
Cable for illumination intensity control
2
HAL 100 light source
3
Clamping screw
4
Illumination mount
5
Stand
6
External power supply for HAL 100 light
source
7
Mains socket
8
Plug of the light source's cable
Parts and Tools
Hex Key, 3.0 mm
Prerequisite
ü
The microscope is switched off.
ü
The stand
5
is equipped with the illumination mount
4
.
ü
The protective cap is removed from the illumination mount.
ü
The base plate is installed. [
ü
The external power supply for HAL 100 light source
6
is switched off.
ü
from the light source housing.
Procedure
1. Loosen the clamping screw
3
on the illumination mount
4
of the carrier for trans-
mitted-light illumination.
2. Insert the dovetail ring of the HAL 100
2
into the illumination mount.
3. Tighten the clamping screw
3
.
4. At the external power supply, insert the plug of the light source's cable
8
into the
TL
port.
5. At the external power supply, insert the cable for illumination intensity control
1
into
the
Remote
port.
6. At the back of the stand, insert the cable for illumination intensity control into the
Remote
port.
7. Switch the toggle switch for reflected or transmitted light to the
TL
position (transmitted
light).
8. Connect the mains socket
7
of the external power supply to a mains. Use the power ca-
ble.
Proceed in the reverse order for removal.
76
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001