6 Care and Maintenance | 6.3 Maintenance Work
ZEISS
6.3.8 Exchanging the Fuses in the Stand
DANGER
Electric shock due to live parts
When the microscope is still switched on, coming in contact with live parts can lead to electric
shock or burn.
4
Switch off the microscope prior to opening or cleaning.
4
Disconnect live parts from the power supply.
2
1
3
1
Fuse
2
Fuse holder
3
Fuse compartment
Parts and Tools
2x Fuse type T3.15A/H/250V
Prerequisite
ü
The microscope is switched off and disconnected from the mains.
Procedure
1. If the fuses fail, first check the cause and remedy technical problems properly.
2. Remove the fuse holder
2
on the rear side of the stand.
3. Remove fuses
1
from the fuse holder.
4. Insert new fuses.
5. Push the fuse holder back into the fuse compartment
3
until it locks in place.
6. Bring the microscope back into operation.
128
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001