7 Trouble-shooting |
7.2 Hardware
User Guide Axiocam 702 mono | V1.0 en
23
7.1.3 The color of my image does not correspond to the
impression through the ocular
In case of color cameras:
¢
Check white balance and, if necessary, repeat white balance.
¢
Check the monitor’s color temperature setting. If necessary, reduce this to the
lowest value that can be set (usually 5200 K).
In case of monochrome cameras:
¢
Images of monochrome cameras are displayed by using overlay colors to
represent the colors of fluorescence dyes on the monitor. If the color is
different to the visual impression in the eyepiece, just select a different overlay
color from the menu in the software (Display tab).
Trouble-shooting |
7.2 Hardware
7.2.1 Vibrations
Sudden jolt, vibration or moving objects during the acquisition can detract the
image quality. The intensity of light during exposure time can change as well.
¢
To avoid jolt use damping systems on your microscope.
¢
Repeat the shot and take care to ensure the conditions remain calm.