6.4.1.1. Preparing the sample holder
IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times. Always wear lint-free gloves when touching specimen, sample holder or
stage.
1
Attach the specimen to the stub by using con-
ductive carbon, adhesive metal or carbon tape
etc.
Ensure that the specimen area to be analysed
is in proper contact with the stub.
2
Use the tweezers to insert the stub into the
sample holder.
3
Properly fix the stub to the sample holder.
Use the Allen wrench to tighten the location
screw.