Cooling water supply
Water flow rate
75 - 85 l/h
Input pressure
0.2 - 0.3 MPa (2 - 3 bar)
Water temperature
20 - 22°C
Stability
0.5°C/10 min
Gas supplies
Nitrogen
Quality
4.6 with nitrogen content <99.996 %
Standard
Flow rate
Approx. 40 l/min for ventilation of specimen
chamber with chamber door open
Pressure
0.30 - 0.35 MPa (3.0 - 3.5 bar)
Compressed air
Flow rate
Less than 1l/min
Pressure
0.6 - 0.8 MPa (6 - 8 bar)
Quality
Oil-free
Environmental requirements
Ambient Temperature
21°C ±4°C
Stability 0.5°C
Relative humidity
Less than 65 %
Altitude
To guarantee undisturbed operation, do not operate the workstation at sites
higher than 2000 m above sea level.
Pollution degree
2
According to EN 61010-1: Safety requirements for electrical equipment for
measurement, control, and laboratory use. Part 1: General requirements.
Содержание AURIGA Compact Crossbeam
Страница 1: ...AURIGA Compact Crossbeam workstation Instruction Manual ...
Страница 50: ...RM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP PK aÉëÅêáéíáçå ìëíçãÉê ëÉêîáÅÉ ...
Страница 54: ...RQ çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP RK fåëí ää íáçå ...
Страница 150: ...NRM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Страница 156: ...NRS çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NNK ÄÄêÉîá íáçåë ...
Страница 160: ...NSM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...