ACCURA Installation site requirements
24
ACCURA 16/42/15 (bellows cover) installation dimensions
640
2627
670
4616
≥
200
2020
1927
1651
3260
2750
1564
5720
4940
÷
All dimensions in millimeters, scale 1:50
600
680
≥
800
≥
800
CMM width
Length of granite plate, approx. 130 mm extra for service
unit.
Length of the CMM with covers
CMM height
Height of secured CMM assembly with shipping pallet.
Height of secured CMM assembly from lower edge of
granite plate.
Height between the granite plate and lower edge of the
stylus mounting cube on the VAST XT gold probe.
With other VAST gold or RDS probe types, the height is
somewhat less.
Network connection for data backup and Internet (Tele-
service)
Power supply
Compressed air supply
TCC (standard) controller cabinet, MCC 800 (option)
➤
see Controller cabinet installation [
18]
CMM <-> controller connection cable (cable length ap-
prox. 3 m)
Compressed air connection on the CMM
Scanner protective field option
Weight:
ACCURA (size 16/42/15)
18540 kg
Workpiece, max.
4000 kg