ACCURA Installation site requirements
14
Electrical Specifications and Network Data
Safety instructions
Electric voltage
WARNING
Danger to life due to electric shock resulting from contact with electrical lines and com-
ponents.
Cardiac arrest, burns, and death.
ü
Any work on the controller must be carried out only by a certified electrician. The cover of
the controller may only be removed when the CMM is out of operation. The following mea-
sures need to be taken for this:
Turn off the drives and the controller.
Turn off the main switch on the controller and secure it against being turned on again uninten-
tionally.
Warning: some components in the controller cabinet remain live even when the main switch is
turned off. These places are marked with warning signs. If necessary, completely disconnect the
controller from the power supply (MCC 800 and MCC 1200, if provided).
If you do not have a permanent connection, pull the power plug as well.
Routing the connecting cables
CAUTION
Tripping hazard due to exposed cables.
Light to moderate bruises and scrapes. Concussions also possible.
Route the cables in a cable conduit or cable duct. The cable conduit is fastened to the floor. The
cable duct is embedded in the floor.
NOTE
The cable conduit is not included with delivery. The cable conduit or cable duct must accommodate
all cables and hoses.