Version 7.0
G-30-1781-en
Page 21
S100 / OPMI pico
Safety Measures
LED light source
– Temperature monitoring
A sensor constantly monitors the LED temperature.
1
LED error indicator
If the amber LED is illuminated, the LED light source is in a limited oper-
ating mode.
In this case, the light intensity drops to 50%.
• Please contact ZEISS Service.
Fig. 3: Safety equipment:LED
light source
CAUTION
LED equipment, laser class 2
The LED light source is a class 2 LED device according to IEC 60825-1:2001.
According to EN 62471:2009 this light source is to be classified as belonging
to the free group.
NOTE
Varying color impression
The color temperature of the lamps differs. This can result in a changing color
impression of the treatment / surgical field.
1
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...