6.4
Cleaning and disinfection
6.4.1 Cleaning
The platform and accessories are cleaned with a moist cloth while the device is switched
off and the mains plug taken out.
NOTE
Do not use any aggressive agents to clean the measuring system.
WARNING
Please be absolutely sure to switch off the device and pull the mains
plug out of the socket before you commence disinfecting and clean-
ing
6.4.2 Manual Disinfection
The platform can be disinfected by wiping over with suitable agents. To clean, wipe the
platform with a cloth soaked in disinfection liquid.
WARNING
No spray disinfection
!
Spray disinfection can destroy the highly precise
measurement sensors of the platform.
Recommended disinfection agent
Composition approx. 25% ethanol, 35% Propanol
E.g. Mikrozid Liquid / Schülke & Mayr or similar agents
NOTE
If you apply disinfection agent be sure to follow the recommenda-
tions given by the manufacturer of the disinfection agent strictly. Es-
pecially consider the rules concerning the commended application
time of the agent.
WARNING
Due to danger of confusion, chemicals that are necessary for the
disinfection or cleansing exclusively must be stored, prepared and
provided in containers that are appropriate for this purpose.
NOTE
In order to demonstrate that disinfection was successfully done, it is
recommended to put up a sign on the platform, saying “disinfected”.